Type N VNA Calbration Kit (DC - 18 GHz) - MTC-31101



Details  
 

Features:

ο Type N50Ω Coaxial Port
ο Broadband: DC-18GHz
ο Fixed Load


General Description:

AV31101 Type N 50Ω coaxial calibration kit is a kind of instrument calibration kit, which can satisfy all demands except measure laboratory. It can be used to calibrate vector network analyzer with type N port. The kit contains the following items:

Serial Number

Model

Name

amount

remark

1

AV81801

Type N male open

1

 

2

AV81802

Type N female open

1

 

3

AV81803

Type N male short

1

 

4

AV81804

Type N female short

1

 

5

AV70501

Type N female load

1

 

6

AV70502

Type N male load

1

 

7

AV71101

N-50KK adapter

1

 

8

AV71102

N-50JJ adapter

1

 

9

AV71103

7mm-type N male adapter

2

 

10

AV71104

7mm-type N female adapter

2

 

11

 

Type N torque wrench

1

 

12

 

Spanner wrench

1

 

Specifications:

1. Electrical Specifications(Temperature: 23±3°C,Frequency Range: DC~18GHz)

Model

Name

Specification

AV81801

Type N male open

Deviation from Nominal Phase
±1.5°

AV81802

Type N female open

AV81803

Type N male short

Deviation from Nominal Phase
±1.0°

AV81804

Type N female short

AV70501

Type N female load

VSWR≤1.055

AV70502

Type N male load

AV71101

N-50KK adapter

VSWR≤1.083

AV71102

N-50JJ adapter

AV71103

7mm-type N male adapter

VSWR≤1.083

AV71104

7mm-type N female adapter

 

Type N torque wrench

Torque Setting 135.6 N.cm±4.1 N.cm

2. Environmental Requirements

Operating temperature: 23°C±3°C
Storage temperature: -20°C to +70°C


Ordering Information:

AV31101

Type N50Ω Coaxial Calibration Kit

Notes:

1. WHEN USING, PLEASE ADOPT ELECTROSTATIC SENSITIVE DEFENSE STEP.

2. BEFORE USING, INSPECT THE CONNECTOR TO MAKE SURE NO PARTICLES OR RESIDUE ARE PRESENT; PLEASE INSURE PIN DEPTH OF DEVICES CONNECTED TO BE UP TO THE STANDARD.

3. WHEN USING, DO NOT TURN THE DEVICE BODY. ONLY TURN THE CONNECTOR NUT. DAMAGE TO THE CENTER CONDUCTOR CAN OCCUR IF THE DEVICE BODY IS TWISTED.

4. USING A TORQUE WRENCH GUARANTEES THAT THE CONNECTION IS NOT TOO TIGHT, PREVENTING POSSIBLE CONNECTOR DAMAGE.

5. WHEN NOT IN USE, PLEASE INSTALL THE PROTECTIVE END CAPS AND STORE THE CALIBRATION DEVICES IN THE FOAM-LINED STORAGE CASE.

6. PLEASE VERIFY THE PERFORMANCE OF YOUR CALIBRATION KIT IN A PERIOD OF TIME BY THE QUALIFIED.

7. PRODUCT SPECIFICATIONS AND DESCRIPTIONS IN THIS DOCUMENT ARE SUBJECT TO CHANGE WITHOUT NOTICE.




  • Share with Us:




  • Company

    Microwave Town LLC is a leader in the microwave component industry. We offer a wide range of high quality, precision RF products such as coaxial and waveguide components, broadband power amplifiers, and antennas for DC to 330 GHz. Our millimeter wave products are carefully designed and widely used in military and commercial applications. It is our focus to meet your stringent specifications and requirements. We are committed to providing our customers with the highest quality products, competitive pricing, fast delivery, and excellent customer service.